Airway ventilation pressures during bronchoscopy, bronchial blocker, and double-lumen endotracheal tube use: an in
Alexander S Kuo1, James H Philip1, Thomas Edrich1
1Department of Anesthesiology, Perioperative and Pain Medicine, Brigham and Women's Hospital, Harvard Medical School, Boston, MA.
Objective:
To quantify inspiratory flow resistance of instrumented single-lumen and double-lumen endotracheal tubes.
Design:
Bench-top in vitro experiments.
Setting:
Laboratory of a university hospital.
Participants:
In vitro lung simulator.
Interventions:
A lung simulator was ventilated mechanically via several single- and double-lumen endotracheal tubes (ETT) that were instrumented with adult and pediatric bronchoscopes as well as bronchial blockers. While ventilating with a square-flow wave and increasing peak inspiratory flow from 10-100 L/min, the pressures proximal and distal to the instrumented ETT were measured. Flow (Q) and the pressure loss (∆P) were related with regression of the quadratic equation: ∆P=k1Q+k2Q2.
Measurements And Main Results:
With all combinations of single-lumen endotracheal tubes, double-lumen endotracheal tubes, bronchial blockers, and adult and pediatric bronchoscopes, ∆P was accurately related to Q using the quadratic equation with excellent fit, R2>0.99 for all combinations. The regression parameters k1 and k2 were statistically significant for all combinations except k1 with a bronchoscope through 37-Fr double-lumen endotracheal tube. Parameter k2 was dominant at flows above 10 L/min for uninstrumented airways and 20 L/min for instrumented airways. ∆P increased dramatically with flow, and increased with decreasing endotracheal tube size or addition of instrumentation in a quantitatively predictable manner.
Conclusions:
Pressure loss across instrumented endotracheal tubes follows a predictable flow-dependant quadratic pattern. Using the quantitative in vitro results of this study, a clinician can maximize inspiratory ventilation pressures during these situations without delivering excessive airway pressures to the patient.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due to...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...