Saliva sampling method affects performance of a salivary α-amylase biosensor
Theodore F Robles1, Rassilee Sharma, Lauren Harrell
1Department of Psychology, University of California, Los Angeles, Los Angeles, 90095-1563, California.
Saliva sampling method significantly impacts salivary alpha-amylase (sAA) biosensor accuracy. Passive drool collection provides more reliable sAA measurements than direct mouth readings for the biosensor.
Area of Science:
- Biochemistry
- Analytical Chemistry
- Biomedical Engineering
Background:
- Salivary alpha-amylase (sAA) is a biomarker used in clinical settings.
- Discrepancies exist between point-of-care biosensors and laboratory analyzers for sAA measurement.
- The method of saliva collection may influence biosensor performance.
Purpose of the Study:
- To investigate how different saliva sampling methods affect the analytical performance of an sAA biosensor.
- To compare direct mouth readings with readings from passively collected saliva.
- To evaluate the impact of sample processing on biosensor accuracy.
Main Methods:
- Compared direct mouth biosensor readings with biosensor and conventional assay readings from passively collected saliva.
- Analyzed saliva samples processed via three methods: unprocessed, thawed, and thawed/centrifuged.
- Utilized Bland-Altman plots to assess agreement between measurement methods.
Main Results:
- Direct mouth readings yielded significantly lower sAA values compared to all other methods.
- Passive drool saliva readings correlated strongly across different processing and assay types.
- Smaller discrepancies were observed between conventional and biosensor readings from passive drool samples.
Conclusions:
- The saliva sampling method critically influences sAA biosensor analytical performance.
- Indirect sAA measurements from passive drool saliva offer more accurate estimations than direct mouth readings.
- Passive drool collection is recommended for reliable sAA biosensor measurements, even with sample processing.