Influence of the hand squeeze and mask distensibility on tidal volume measurements during neonatal mask ventilation
Jeroen J van Vonderen1, Rita Kamar, Kim Schilleman
1Department of Pediatrics, Division of Neonatology, Leiden University Medical Center, Leiden, The Netherlands.
Background:
During mask ventilation, the mask volume can vary as it is pressurized or when it is squeezed. The change in volume of the mask may affect tidal volumes delivered and difference in inspired (Vti) and expired tidal volumes (Vte).
Objectives:
To investigate whether hand squeeze and distensibility of the mask during ventilation influences tidal volume measurements.
Methods:
For both experiments, we ventilated a leak-free mask ventilation model using pressures of 25/5 cm H2O through a t-piece. Vti and Vte were measured. (A) Two consultants performed mask ventilation with (1) consistent hand squeeze, (2) release during inflation and squeeze during expiration, (3) squeeze during inflation, release during expiration, and (4) gentle squeeze.
Results:
(B) Thirty caregivers performed mask ventilation. Experiment A: Vti was different during consistent hold (1) 8.1 ml (0.4) and loose grip (4) 8.2 ml (0.3), compared to squeezing during inflation (2) 18.9 ml (1.9), or expiration (3) 6.4 ml (3.5). Variance in difference between Vti and Vte occurred only when the mask was squeezed during inflation (-47.4% (101.5)). Experiment B: volumes measured were consistent (intraindividual CV 3-5%, interindividual CV 9-10%). When comparing gas flow rate of 6-10 l/min, volumes increased by approximately 8%, differences in Vti and Vte were small with both flow settings (-0.9% (-3.9-1.4) and -0.6% (-3.3-1.8); n.s.).
Conclusion:
Variation in mask hold during mask ventilation can influence volume measurement, but this hardly occurs when testing caregivers.
Related Concept Videos
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...