Precision of In Vivo Quantitative Tooth Wear Measurement Using Intra-Oral Scans
Published on: July 12, 2022
Quantitative evaluation of dental abfraction and attrition using a swept-source optical coherence tomography system
Corina Marcauteanu1, Adrian Bradu, Cosmin Sinescu
1Universitatea de Medicina si Farmacie "Victor Babes," Facultatea de Medicina Dentara, Timisoara 300041, Romania.
Abstract:
A fast swept-source optical coherence tomography (SS-OCT) system is employed to acquire volumes of dental tissue, in order to monitor the temporal evolution of dental wear. An imaging method is developed to evaluate the volume of tissue lost in ex vivo artificially induced abfractions and attritions. The minimal volume (measured in air) that our system could measure is 2352 μm3. A volume of 25,000 A-scans is collected in 2.5 s. All these recommend the SS-OCT method as a valuable tool for dynamic evaluation of the abfraction and attrition with remarkable potential for clinical use.