Related Experiment Video
Updated: May 7, 2026

Detection of 3-Nitrotyrosine in Atmospheric Environments via a High-performance Liquid Chromatography-electrochemical Detector System
Published on: January 30, 2019
[Method for determining o-tolidine in workplace air by gas chromatography]
He-xia Zha1, Hao Zhao, Shi-quan Zhao
1Nanjing Center for Disease Control and Prevention, Nanjing, Jiangsu 210003, China.
Objective:
To establish a method for determining o-tolidine in workplace air by gas chromatography.
Methods:
o-tolidine in workplace air was collected with a glass fiber filter, desorbed with methanol, and determined by gas chromatography-flame ionization detector.
Results:
The concentration of o-tolidine showed a linear relationship with peak area within 0.04∼9.00 µg/ml; the detection limit was 0.04 µg/ml; the minimum detectable concentration was 0.0002 mg/m(3) (calculated by 375 L air sample); the sampling efficiency was 93%∼100%; the elution efficiency was 94%∼96%; the relative standard deviation was 0.8-2.5%. Sample could be stored at 4 °C for at least 8 days and at room temperature for as long as 6 days.
Conclusion:
This determination method meets the requirements of Guide for establishing occupational heath standards-Part 4 Determination methods of air chemicals in workplace (GBZ/T 210.4-2008) and is suitable for determination of o-tolidine in workplace air.
More Related Videos
08:23Analyzing the Photo-oxidation of 2-propanol at Indoor Air Level Concentrations Using Field Asymmetric Ion Mobility Spectrometry
Published on: June 14, 2018
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Related Concept Videos
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...