Dose distribution response in HDRB measured with EBT2 and compared with PLATO SYSTEM
L Hernández-Ruiz1, J O Hernández-Oviedo, D Ruesga-Vazquez
1Centro de Investigación en Ciencia Aplicada y Tecnología Avanzada-Legaria del Instituto Politécnico Nacional, Av. Legaria 694, 11500 México City, Mexico.
High Dose Rate Brachytherapy (BHDR) dose distribution using Gafchromic film was validated against treatment planning. Results show good agreement, with a minor 2.11% difference between planned and measured isodoses for (192)Ir treatments.
Area of Science:
- Medical Physics
- Radiotherapy Dosimetry
- Radiation Oncology
Background:
- Accurate dose distribution is critical for effective High Dose Rate Brachytherapy (BHDR).
- Gafchromic films are used for dosimetric measurements in radiotherapy.
- The energy dependence of Gafchromic film response needs careful calibration.
Purpose of the Study:
- To measure and validate the dose distribution of (192)Ir High Dose Rate Brachytherapy (BHDR).
- To assess the accuracy of the PLATO treatment planning system by comparing planned and measured isodoses.
- To investigate the energy dependence of Gafchromic EBT2 film response.
Main Methods:
- Dose distribution was measured using Gafchromic EBT2 film for (192)Ir HDR BHDR.
- Film calibration was performed using (60)Co and 6/18 MV LINAC units.
- Experimental dose distributions were compared with those generated by the PLATO planning system.
Main Results:
- Experimental dose distribution measurements closely matched the PLATO treatment planning system.
- A mean difference of 2.11% was observed between the planned and measured isodoses.
- The study confirmed the feasibility of using Gafchromic film for HDR BHDR dosimetry.
Conclusions:
- Gafchromic EBT2 film is a reliable dosimeter for (192)Ir HDR BHDR.
- The PLATO planning system provides accurate dose predictions for HDR BHDR.
- Precise calibration is essential due to the energy-dependent response of Gafchromic films.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Measurement of Bioavailability: Pharmacodynamic Methods
Measurement of Bioavailability: Pharmacokinetic Methods
Pharmacokinetic–Pharmacodynamic Relationship: Dose to Pharmacological Effect
Pharmacokinetic–Pharmacodynamic Relationship: Problems
Pharmacokinetic–Pharmacodynamic Relationship: Intensity of Dose-Effect Relationship
Compartment Models: Two-Compartment Model