Related Experiment Video
Updated: May 7, 2026

Measurement of Healthy and Injured Triceps Surae Morphology
Published on: October 27, 2023
Test-retest reliability and the minimal detectable change for achilles tendon length: a panoramic ultrasound
Eric D Ryan1, Joseph G Rosenberg, Michael J Scharville
1Neuromuscular Research Laboratory, Department of Exercise and Sport Science, University of North Carolina-Chapel Hill, Chapel Hill, North Carolina, USA.
Abstract:
The purpose of the present study was to examine test-retest reliability and minimal detectable change (MDC) values for Achilles tendon (AT) length determined using panoramic ultrasound (US) imaging. Seventeen men (age = 21.0 ± 2.3 y) visited the laboratory on two separate days, where AT length was examined along the mid-longitudinal axis of the right lower leg with a portable B-mode panoramic US device. These measures were found to have acceptable reliability with intra-class correlation coefficient (ICC) and standard error of measurement (SEM) values of 0.954 and 4.43 mm (SEM% of the mean = 2.37%), respectively. In addition, the MDC for the panoramic US assessment of AT length was 12.27 mm (MDC% of the mean = 6.57%). These findings suggest that panoramic US imaging is a reliable technique for detection of clinically relevant changes in AT length and may therefore be a practical and time-efficient clinical tool for future studies examining AT length in vivo.

