Effects of Surgical Masks on Cardiopulmonary Function in Healthy Subjects
Published on: February 12, 2021
[Protective performance and subjective evaluation of N95 filtering-facepiece respirators]
Lu-man Jiang1, Yan-yan Yu, Yi Rong
1School of Public Health, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430030, China.
Objective:
To evaluate the protective performance of N95 filtering-facepiece respirators (FFRs) used widely in China and to investigate participants' subjective evaluation about them.
Methods:
Four models (A1, A2, B1, B2) of N95 FFRs from two manufactures were chosen to measure the filter penetration and inhalation resistance. Inward leakage was measured by Condensation Nuclei Counting method (CNC) in 50 participants selected using the Chinese respirator fit test panel. Each participant was asked subjective feelings after wearing a respirator by questionnaire survey.
Results:
The filter penetration and inhalation resistance of four FFRs complied with national standard (GB 2626-2006). The geometric mean fit factors (GMFFs) for four models were 20.9, 14.6, 74.0, 49.1 and there passing rates were 4%, 4%, 42%, 10%. All of four models had bad seal performance, especially the passing rate of A1 and A2 were lower than 10%. The self-feelings about the resistance for FFRs had no significant difference (P > 0.05). The results indicated that B (B1 and B2) has a better fit than A (A1 and A2) according to participants' subject evaluation (P < 0.05). The folding style respirator (B2) was significant more comfortable than the cup style respirator (B1). The subjective feeling of respirators leakage by participant was poor consistent with objective inward leakage test. The kappa index was 0.067 (95%CI: -0.029∼0.163, P = 0.18) and the consistent rate was 50%.
Conclusion:
The poor seal performance was the biggest problem of N95 FFRs in Chinese market. Respirators should be resigned or improved rely on Chinese facial features. Dust workers should choice a fit respirator according to the result of objective leakage test rather than subjective feeling of leakage.
Related Concept Videos
Personal Protective Equipment
PPE Use in Healthcare Settings I: Donning
PPE Use in Healthcare Settings II: Doffing
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask and...