Related Experiment Video
Updated: May 7, 2026

Multispectral Real-time Fluorescence Imaging for Intraoperative Detection of the Sentinel Lymph Node in Gynecologic Oncology
Published on: October 20, 2010
Transferring innovative freehand SPECT to the operating room: first experiences with sentinel lymph node biopsy in
A L Mihaljevic1, A Rieger1, B Belloni2
1Department of Surgery, Klinikum rechts der Isar, Technische Universität München, Ismaningerstrasse 22, 81675 Munich, Germany.
Purpose:
The purpose of this study was to report on the first experiences with freehand single-photon emission-computed tomography (freehand SPECT) in sentinel lymph node biopsy (SLNB) in patients with malignant melanoma. Freehand SPECT is a novel imaging modality combining gamma probes, surgical navigation systems, and emission tomography algorithms, designed to overcome some of the limitations of conventional gamma probes.
Methods:
In this study 20 patients with malignant melanoma underwent conventional planar scintigraphy prior to surgery. In the operating room, the number and location of separable SLNs were detected first by a pre-incisional scan with freehand SPECT to render a 3D-image of the target site and afterwards by a scan with a conventional gamma probe. After SLNB another scan was performed to document the removal of all targeted SLNs.
Results:
Planar scintigraphy identified 40 SLNs in 26 nodal basins. Pre-incisional freehand SPECT mapped 38 of these nodes as well as one additional node in one patient (95.0% node based sensitivity). The results of freehand SPECT were identical to those of planar scintigraphy in 25 basins, while it missed one basin (96.2% basin based sensitivity). In comparison, the gamma probe failed to detect 7 nodes in 4 basins (82.5% node based sensitivity and 84.6% basin based sensitivity). After resection freehand SPECT detected 9 remaining radioactive spots, two of whichwere resected as they matched the position of SLNs detected on preoperative planar scintigraphy.
Conclusions:
Freehand SPECT provides a real-time, intraoperative 3D-image of the radioactive labelled SLNs, facilitating their detection and resection.