Field camera measurements of gradient and shim impulse responses using frequency sweeps
S Johanna Vannesjo1, Benjamin E Dietrich, Matteo Pavan
1Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland.
Purpose:
Applications of dynamic shimming require high field fidelity, and characterizing the shim field dynamics is therefore necessary. Modeling the system as linear and time-invariant, the purpose of this work was to measure the impulse response function with optimal sensitivity.
Theory And Methods:
Frequency-swept pulses as inputs are analyzed theoretically, showing that the sweep speed is a key factor for the measurement sensitivity. By adjusting the sweep speed it is possible to achieve any prescribed noise profile in the measured system response. Impulse response functions were obtained for the third-order shim system of a 7 Tesla whole-body MR scanner. Measurements of the shim fields were done with a dynamic field camera, yielding also cross-term responses.
Results:
The measured shim impulse response functions revealed system characteristics such as response bandwidth, eddy currents and specific resonances, possibly of mechanical origin. Field predictions based on the shim characterization were shown to agree well with directly measured fields, also in the cross-terms.
Conclusion:
Frequency sweeps provide a flexible tool for shim or gradient system characterization. This may prove useful for applications involving dynamic shimming by yielding accurate estimates of the shim fields and a basis for setting shim pre-emphasis.
More Related Videos
14:23Optical Recording of Electrical Activity in Guinea-pig Enteric Networks using Voltage-sensitive Dyes
Published on: December 4, 2009
09:01Gain-compensation Methodology for a Sinusoidal Scan of a Galvanometer Mirror in Proportional-Integral-Differential Control Using Pre-emphasis Techniques
Published on: April 4, 2017
Related Concept Videos
Impulse Response
Kirchhoff's law forms an input signal equation, with the capacitor's current and voltage providing the output. Substituting the current and dividing by RC yields a differential equation. The output for an impulse input is the impulse...
Reconstruction of Signal using Interpolation