Related Experiment Video
Updated: May 7, 2026

Modeling and Experimental Analysis of the Single-Shaft Coaxial Motor-Pump Assembly in Electrohydrostatic Actuators
Published on: June 13, 2022
Effect of impeller flow path on pump performance and impeller stability of the monopivot circulatory pump
Abstract:
The effect of a cutout on the pump pressure-flow characteristics and the impeller stability was quantified using computational fluid dynamics analysis in order to provide good hemocompatibility of the monopivot extracorporeal circulation pump. As a result, the following findings were clarified. The pump pressure is lower in the cutout model than in the no-cutout model. The impeller stability with respect to the buoyancy of the impeller is better in the cutout model than in the no-cutout model. The impeller stability with respect to the impeller tilt is better in the cutout model than in the no-cutout model. Therefore, the cutout model, in which the geometry corresponds to the commercialized pump, was likely to be better than the no-cutout model because the stability that has the possibility to decrease the gap instantaneously to increase hemolysis despite the impeller rotational speed slightly.
Related Concept Videos
Laminar Flow
Steady, Laminar Flow Between Parallel Plates
Steady, Laminar Flow in Circular Tubes
Typical Model Studies
Steady Flow of a Fluid Stream
During this process, the momentum of the fluid within the control volume remains constant over the time interval dt. By applying the...