Related Experiment Video
Updated: May 7, 2026

05:25
Identification and Protection of the Recurrent Laryngeal Nerve during Transoral Robotic Thyroidectomy
Published on: October 24, 2025
964
Robotic microlaryngeal surgery: feasibility using a newly designed retractor and instrumentation
Ian J Lalich1, Steven M Olsen, Dale C Ekbom
1Department of Otorhinolaryngology, Mayo Clinic, Rochester, Minnesota, U.S.A.
The Laryngoscope
|October 12, 2013
Summary
Researchers developed new robotic retractors and instruments for robotic microlaryngeal surgery (RMLS). This feasibility study in cadavers shows potential for improved glottic exposure and delicate surgical procedures.
Area of Science:
- Robotic Surgery
- Head and Neck Surgery
- Laryngeal Surgery
Background:
- Robotic surgery is expanding in head and neck procedures but not yet routine for microlaryngeal surgery.
- Current retractors limit tongue and oral commissure retraction, larynx visualization, and robotic adaptability.
- Existing microlaryngeal surgery instrumentation is inadequate for robotic applications.
Purpose of the Study:
- To assess the feasibility of robotic microlaryngeal surgery (RMLS) using a novel robotic retractor.
- To adapt existing microlaryngeal instrumentation for robotic use.
- To achieve routine glottic exposure for RMLS.
Main Methods:
- A feasibility study was conducted using a newly designed robotic retractor in a cadaver model.
- Microlaryngeal instruments were adapted for the da Vinci Surgical Robot.
- The study aimed to evaluate glottic exposure and robotic instrument access.
Main Results:
- Robotic microlaryngeal surgery was successfully performed in a cadaver model.
- Procedures included microflap elevation, anterior commissure surgery, cordectomy, and arytenoidectomy.
- Photographic and video documentation confirmed the feasibility of the techniques.
Conclusions:
- A novel laryngeal robotic retractor and microlaryngeal instrumentation were developed and tested in a cadaver model.
- The system allows for glottic and anterior commissure exposure and robotic access for delicate endolaryngeal procedures.
- Further evaluation in an in vivo human model is required to determine the retractor's efficacy.

