Lateral temperature spread of monopolar, bipolar and ultrasonic instruments for robot-assisted laparoscopic surgery
Lukas J Hefermehl1, Remo A Largo, Thomas Hermanns
1Division of Urology, University Hospital Zurich, Zurich, Switzerland.
Objective:
To assess critical heat spread of cautery instruments used in robot-assisted laparoscopic (RAL) surgery.
Materials And Methods:
Thermal spread along bovine musculofascial tissues was examined by infrared camera, histology and enzyme assay. Currently used monopolar, bipolar and ultrasonic laparoscopic instruments were investigated at various power settings and application times. The efficacy of using an additional Maryland clamp as a heat sink was evaluated. A temperature of 45 °C was considered the threshold temperature for possible nerve damage.
Results:
Monopolar instruments exhibited a mean (sem) critical thermal spread of 3.5 (2.3) mm when applied at 60 W for 1 s. After 2 s, the spread was >20 mm. For adjustable bipolar instruments the mean (sem) critical thermal spread was 2.2 (0.6) mm at 60 W and 1 s, and 3.6 (1.3) mm at 2 s. The PK and LigaSure forceps had mean (sem) critical thermal spreads of 3.9 (0.8) and 2.8 (0.6) mm respectively, whereas the ultrasonic instrument reached 2.9 (0.8) mm. Application of an additional Maryland clamp as a heat sink, significantly reduced the thermal spread. Histomorphometric analyses and enzyme assay supported these findings.
Conclusions:
All coagulation devices used in RAL surgery have distinct thermal spreads depending on power setting and application time. Cautery may be of concern due to lateral temperature spread, causing potential damage to sensitive structures including nerves. Our results provide surgeons with a resource for educated decision-making when using coagulation devices during robotic procedures.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...