Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Optical quality of hyperopic and myopic phakic intraocular lenses
Cari Pérez-Vives1, Alberto Domínguez-Vicent, Teresa Ferrer-Blasco
1Department of Optics, Faculty of Physics, University of Valencia, Valencia, Spain.
Aims:
To assess and compare the optical quality of the myopic and hyperopic implantable collamer lens (ICL) from its wavefront aberrations for different powers and pupil diameters.
Settings And Design:
Prospective study.
Material And Methods:
The wavefront aberrations of two myopic (-3 and -6 diopters (D)) and two hyperopic V4b ICLs (+3 and +6D) were measured in vitro. To assess and compare the optical quality of different powers of ICLs, we analyzed the root mean square (RMS) of total higher order aberrations (HOAs), trefoil, coma, tetrafoil, secondary astigmatism, and spherical aberration at 3- and 4.5-mm pupil. In addition, the point spread functions (PSFs) of each ICL evaluated were calculated from the wavefront aberrations at 3- and 4.5-mm pupil.
Statistical Analysis:
A Student's t-test for unpaired data was used for comparison between myopic and hyperopic ICLs.
Results:
Myopic ICLs showed negative spherical aberration, in contrast hyperopic ICLs showed positive spherical aberration, which increases when the ICL power increases, due to the innate optical properties of the lens. All ICLs evaluated had negligible amounts of other aberrations. We did not find statistical significant differences in any Zernike coefficient RMS values analyzed between myopic and hyperopic ICLs at 3- and 4.5-mm pupil (P > 0.05).
Conclusions:
Myopic and hyperopic ICLs provide good and comparable optical quality for low to moderate refractive error. The ICLs evaluated showed values of wavefront aberrations clinically negligible to affect the visual quality after implantation.
More Related Videos
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
Related Concept Videos
Focusing of Light in the Eye
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Glaucoma: Overview
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...