Related Experiment Video
Updated: May 6, 2026

Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Repeatability of corneal epithelial thickness measurements using Fourier-domain optical coherence tomography in
Xingxuan Jack Ma1, Li Wang, Douglas D Koch
1Department of Ophthalmology, Cullen Eye Institute, Baylor College of Medicine, Houston, TX.
Purpose:
To evaluate the repeatability of corneal epithelial thickness (ET) and corneal thickness (CT) measurements in normal eyes and laser in situ keratomileusis (LASIK)-treated eyes using optical coherence tomography (RTVue system).
Methods:
In 35 right eyes of 35 normal subjects and 45 right eyes of 45 subjects who underwent a myopic LASIK, corneal ET and CT were evaluated in 17 areas: (1) 1 central zone within a 0- to 2.0-mm diameter, (2) 8 paracentral zones from a 2.0- to 5.0-mm diameter, and (3) 8 peripheral zones from a 5.0- to 6.0-mm diameter. The repeatability was assessed using within-subject standard deviation (SD), coefficient of variation, and intraclass correlation coefficient.
Results:
At the central and paracentral zones, respectively, the SD values were 0.7 μm and 0.6 to 0.9 μm in normal eyes and 0.7 μm and 0.8 to 1.7 μm in LASIK-treated eyes for ET, and 1.0 μm and 2.8 to 4.6 μm in normal eyes and 1.3 μm and 4.0 to 4.8 μm in LASIK-treated eyes for CT. At the peripheral zones, in normal and LASIK-treated eyes, respectively, the SD values ranged from 0.8 to 1.2 μm and 1.4 to 2.2 μm for ET, and 4.1 to 6.4 μm and 6.0 to 9.1 μm for CT. The coefficient of variation values were low and intraclass correlation coefficient values were high in both groups for both ET and CT measurements.
Conclusions:
The Optical coherence tomography produced excellent repeatability, especially at the central and paracentral zones up to a 5-mm diameter for both corneal ET and CT measurements.

