Rapid isotropic resolution cartilage assessment using radial alternating repetition time balanced steady-state
Habib Al saleh1, Larry Hernandez, Kenneth S Lee
1Department of Medical Physics, University of Wisconsin, Madison, Wisconsin, USA.
Purpose:
To compare a balanced steady-state free-precession sequence with a radial k-space trajectory and alternating repetition time fat suppression (Radial-ATR) with other currently used fat-suppressed 3D sequences for evaluating the articular cartilage of the knee joint at 3.0T.
Materials And Methods:
Radial-ATR, fast spin-echo (FSE-Cube), gradient recall-echo acquired in the steady-state (GRASS), and spoiled gradient recall-echo (SPGR) sequences with similar voxel volumes and identical scan times were performed at 3.0T on both knee joints of five volunteers. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) measurements were performed for all sequences using a double acquisition method and compared using Mann-Whitney Wilcoxon tests. Radial-ATR sequences with 0.3 mm and 0.4 mm isotropic resolution were also performed on the knee joints of seven volunteers and three patients with osteoarthritis.
Results:
Average SNR values for cartilage, synovial fluid, and bone marrow were 54.7, 153.3, and 12.9, respectively, for Radial ATR, 30.8, 44.1, and 1.9, respectively, for FSE-Cube, 13.3, 46.9, and 3.3, respectively, for GRASS, and 19.1, 8.1, and 2.1, respectively, for SPGR. Average CNR values between cartilage and synovial fluid and between cartilage and bone marrow were 98.6 and 41.8, respectively, for VIPR-ATR, 13.4 and 28.8, respectively, for FSE-Cube, 33.6 and 10.0, respectively, for GRASS, and 11.0 and 16.9, respectively, for SPGR. Radial-ATR had significantly higher (P < 0.001) cartilage, synovial fluid, and bone marrow SNR and significantly higher (P < 0.01) CNR between cartilage and synovial fluid and between cartilage and bone marrow than FSE-Cube, GRASS, and SPGR. Radial-ATR provided excellent visualization of articular cartilage at high isotropic resolution with no image degradation due to off-resonance banding artifacts.
Conclusion:
Radial-ATR had superior SNR efficiency to other fat-suppressed 3D cartilage imaging sequences and produced high isotropic resolution images of the knee joint which could be used for evaluating articular cartilage at 3.0T.
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies III: Computed Tomography
Magnetic Resonance Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...