Related Experiment Video
Updated: May 6, 2026

Multimodal Volumetric Retinal Imaging by Oblique Scanning Laser Ophthalmoscopy oSLO and Optical Coherence Tomography OCT
Published on: August 4, 2018
Maximum-likelihood estimation in Optical Coherence Tomography in the context of the tear film dynamics
Jinxin Huang1, Eric Clarkson, Matthew Kupinski
1Department of Physics and Astronomy, University of Rochester, Rochester, New York 14627, USA.
Abstract:
Understanding tear film dynamics is a prerequisite for advancing the management of Dry Eye Disease (DED). In this paper, we discuss the use of optical coherence tomography (OCT) and statistical decision theory to analyze the tear film dynamics of a digital phantom. We implement a maximum-likelihood (ML) estimator to interpret OCT data based on mathematical models of Fourier-Domain OCT and the tear film. With the methodology of task-based assessment, we quantify the tradeoffs among key imaging system parameters. We find, on the assumption that the broadband light source is characterized by circular Gaussian statistics, ML estimates of 40 nm +/- 4 nm for an axial resolution of 1 μm and an integration time of 5 μs. Finally, the estimator is validated with a digital phantom of tear film dynamics, which reveals estimates of nanometer precision.