Related Experiment Video
Updated: May 6, 2026

Reverse Total Shoulder Arthroplasty
Published on: July 5, 2011
Outcomes of four-corner arthrodesis using the Hubcap circular plate
Sameer K Khan1, Syed M Ali, Andrew McKee
1Department of Trauma and Orthopaedics, Peterborough & Stamford Hospitals NHS Foundation Trust, Peterborough City Hospital (Edith Cavell Campus), Bretton Gate, Peterborough, PE3 9GZ, UK.
Abstract:
We present results of four-corner carpal arthrodesis with the Acumed Hubcap circular plate performed at our unit. Eight patients underwent eight procedures over five years, for scapholunate advanced collapse (five wrists) and scaphoid non-union advanced collapse (three wrists). Outcomes included range of motion, quickDASH scores, and visual analogue scores for satisfaction. At final follow-up, mean flexion-extension arc was 56°, mean radial-ulnar deviation 29° and mean quickDASH score was 23/100. Mean score for satisfaction was 7.7/10 (77%). Seven out of eight (87.5%) patients said they would have it done again, and would also recommend it to others. Radiological union was achieved in all cases. One screw broke in one arthrodesis without causing symptoms. The functional outcomes with our use of the Hubcap are comparable to those reported in literature to date with other circular plates (e.g. Spider plate). There were no non-unions, which is the main reported complication with these plates.