Response in electrostatic analyzers due to backscattered electrons: case study analysis with the Juno Jovian Auroral
G Clark1, F Allegrini, B M Randol
1Department of Physics and Astronomy, University of Texas at San Antonio, San Antonio, Texas 78249, USA.
Abstract:
In this study, we introduce a model to characterize electron scattering in an electrostatic analyzer. We show that electrons between 0.5 and 30 keV scatter from internal surfaces to produce a response up to ~20% of the ideal, unscattered response. We compare our model results to laboratory data from the Jovian Auroral Distribution Experiment-Electron sensor onboard the NASA Juno mission. Our model reproduces the measured energy-angle response of the instrument well. Understanding and quantifying this scattering process is beneficial to the analysis of scientific data as well as future instrument optimization.
Related Concept Videos
Mass Analyzers: Common Types
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Emission Spectroscopy: Instrumentation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Emission Spectroscopy: Overview