Simultaneous measurement of phenylalanine and tyrosine by high performance liquid chromatography (HPLC) with
Gabriele Neurauter1, Sabine Scholl-Bürgi, Astrid Haara
1Division of Biological Chemistry, Biocenter, Medical University, Innsbruck, Austria.
Objectives:
An HPLC method was developed to quantify serum concentrations of phenylalanine and tyrosine simultaneously using fluorescence detection without derivatization.
Methods:
Serum protein is precipitated with trichloroacetic acid, 0.015mM dihydrogen-phosphate solution is used for separation on reversed-phase C18 material, and acetonitrile is avoided. Both amino acids are monitored utilizing their natural fluorescence at 210nm excitation and 302nm emission wavelengths.
Results:
One analytical run is completed within 7min. Lower detection limit for Phe and Tyr is 0.3μM. Comparison of the new method with a classical HPLC method for total amino acids and using UV-absorption detection reveals a highly significant relationship for Phe and Tyr.
Conclusion:
The new HPLC method allows rapid and very sensitive measurement of phenylalanine and tyrosine concentrations.
More Related Videos
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
10:30A Generalized Method for Determining Free Soluble Phenolic Acid Composition and Antioxidant Capacity of Cereals and Legumes
Published on: June 10, 2022
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation