Measurement of skin dose from cone-beam computed tomography imaging
Sercan Akyalcin1, Jeryl D English, Kenneth M Abramovitch
1Department of Orthodontics, The University of Texas Health Science Center at Houston, School of Dentistry, Houston, TX, USA. Sercan.akyalcin@uth.tmc.edu.
Objective:
To measure surface skin dose from various cone-beam computed tomography (CBCT) scanners using point-dosimeters.
Materials & Methods:
A head anthropomorphic phantom was used with nanoDOT optically stimulated luminescence (OSL) dosimeters (Landauer Corp., Glenwood, IL) attached to various anatomic landmarks. The phantom was scanned using multiple exposure protocols for craniofacial evaluations in three different CBCT units and a conventional x-ray imaging system. The dosimeters were calibrated for each of the scan protocols on the different imaging systems. Peak skin dose and surface doses at the eye lens, thyroid, submandibular and parotid gland levels were measured.
Results:
The measured skin doses ranged from 0.09 to 4.62 mGy depending on dosimeter positions and imaging systems. The average surface doses to the lens locations were ~4.0 mGy, well below the threshold for cataractogenesis (500 mGy). The results changed accordingly with x-ray tube output (mAs and kV) and also were sensitive to scan field of view (SFOV). As compared to the conventional panoramic and cephalometric imaging system, doses from all three CBCT systems were at least an order of magnitude higher.
Conclusions:
Peak skin dose and surface doses at the eye lens, thyroid, and salivary gland levels measured from the CBCT imaging systems were lower than the thresholds to induce deterministic effects. However, our findings do not justify the routine use of CBCT imaging in orthodontics considering the lifetime-attributable risk to the individual.