Some limitations of the cell-attached patch clamp technique: a two-electrode analysis.

Summary

This study examined the limitations of the cell-attached patch clamp technique using a two-electrode setup on chick embryo cardiac cells. The researchers measured seal resistance, patch resistance, channel conductance, and membrane potential. They found that apparent seal resistance may be inaccurate due to the parallel combination of seal and patch resistance. Patch resistance was also found to be lower than typically assumed and influenced by the electrode filling solution. In small cell preparations, single-channel conductance measurements may be inaccurate because membrane potential does not remain constant as electrode potential is varied. The study highlights the importance of accounting for these factors to improve the accuracy of patch-clamp experiments.