Fourier transform ion cyclotron resonance detection of multiphoton ionization spectroscopy
1Department of Chemistry, University of California, Santa Barbara, California, USA.
Abstract:
Fourier transform ion cyclotron resonance (FT-ICR) detection was tested for resonanceenhanced multiphoton ionization (REMPI) spectroscopy. The (2+1) REMPI spectra of acetaldehyde were obtained in the wavelength range 364-354 nm via a two-photon resonant 3s ← n Rydberg transition. The space-charge effects on the REMPI spectra were examined in the vicinity of the 0 0 (0) transition. The trapping efficiency measurement shows that all the ions produced from REMPI dissociation processes are arrested in the ion cyclotron resonance cell even in the presence of space-charge interactions. Axial kinetic energy release distributions of ions were extracted from the trapping efficiency data obtained under a new space-charge-free condition. FT-ICR peak heights were measured as a function of pressure at different laser powers, magnetic field strengths, and ion excitation methods to test for the detection linearity. The FT-ICR detection responds linearly to the number of ions in a low pressure limit. The product branching ratio was measured by using various ion excitation methods and was compared with the previous quadrupole mass spectrometric study. FT-ICR detection yields the mass-selected REMPI spectra and the product branching ratio in the absence of kinetic shifts.
Related Concept Videos
Mass Analyzers: Common Types
Mass Spectrum: Interpretation
Mass Spectrometers
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
IR Spectrometers