Related Experiment Video
Updated: May 6, 2026

Bacterial Detection & Identification Using Electrochemical Sensors
Published on: April 23, 2013
Coupling electrochemical response of a DNA biosensor with PCR for Neisseria gonorrhoeae detection
Rachna Verma1, Seema Sood1, Renu Singh2
1Department of Microbiology, All India Institute of Medical Sciences, New Delhi, India.
Abstract:
Early diagnosis of gonococcal infections is important with regard to a patient's health and stage of infection. In this context, we report the development of an opa-gene-based electrochemical DNA biosensor for detection of Neisseria gonorrhoeae by monitoring redox peak of methylene blue indicator. The fabricated biosensor has been shown to be highly sensitive and specific when evaluated with complementary, non-complementary, and 1-base mismatch DNA sequences and polymerase chain reaction (PCR) amplified products (amplicons) of standard strain of N. gonorrhoeae (ATCC49226). The biosensor has been further evaluated using amplicons of known positive and negative clinical samples, and cut-off for positives has been determined using receiver operating characteristic curve. The sensitivity (SN), specificity (SP), positive predictive value, and negative predictive value of the biosensor have been found to be 96.2%, 88.2%, 92.6%, and 93.8%, respectively. We conclude that the combination of PCR amplification with electrochemical detection shows distinct advantage of high SN and increased SP for gonococcal detection.