Related Experiment Video
Updated: May 6, 2026

Author Spotlight: Development of a Novel Finite Element Analysis Model for Improved Orthognathic Surgical Techniques
Published on: October 20, 2023
Biomechanical optimization of the diameter of distraction screw in distraction implant by three-dimensional finite
Songhe Lu1, Tao Li, Yongqiang Zhang
1Department of Oral and Maxillofacial Surgery, School of Stomatology, The Fourth Military Medical University, Xi'an 710032, PR China; Department of Epidemiology, The Fourth Military Medical University, 169 West Changle Road, Xi'an, Shaanxi 710032, PR China.
Abstract:
Three-dimensional (3D) finite element models of a posterior mandibular segment and a distraction implant (DI) were created and assembled in this study. The diameter of distraction screw (DS) was set as input variable, ranging from 1.0 to 3.0mm, to analyze the stress and displacement, in order to obtain an optimal outcome of the DI. The results indicate that when the diameter of DS was 2.0mm, the stresses in jaw bone and DS and the displacement of DS reached a relatively low range. The diameter of DS significantly affects the stress distribution in DI and surrounding bone. The present study demonstrates that from clinical and biomechanical points of view, the optimal diameter of DS is 2.0mm.