Low-dose computed tomography of the lumbar spine: a phantom study on imaging parameters and image quality
Muhammed Alshamari1, Mats Geijer2, Eva Norrman3
1Department of Radiology, Örebro University Hospital and Department of Health and Medical Sciences, Örebro University, Sweden muhammed.alshamari@orebroll.se.
Background:
Lumbar spine radiography has limited diagnostic value but low radiation dose compared with computed tomography (CT). The average effective radiation dose from lumbar spine radiography is about 1.1 mSv. Low-dose lumbar spine CT may be an alternative to increase the diagnostic value at low radiation dose, around 1 mSv.
Purpose:
To determine the optimal settings for low-dose lumbar spine CT simultaneously aiming for the highest diagnostic image quality possible.
Material And Methods:
An ovine lower thoracic and lumbar spine phantom, with all soft tissues around the vertebrae preserved except the skin, was placed in a 20 L plastic container filled with water. The phantom was scanned repeatedly with various technical settings; different tube potential, reference mAs, and with different convolution filters. Five radiologists evaluated the image quality according to a modification of the European guidelines for multislice computed tomography (MSCT) quality criteria for lumbar spine CT 2004. In a visual comparison the different scans were also ranked subjectively according to perceived image quality. Image noise and contrast were measured.
Results:
A tube potential of 120 kV with reference mAs 30 and medium or medium smooth convolution filter gave the best image quality at a sub-millisievert dose level, i.e. with an effective dose comparable to that from lumbar spine radiography.
Conclusion:
Low-dose lumbar spine CT thus opens a possibility to substitute lumbar spine radiography with CT without obvious increase in radiation dose.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
08:41Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT