Use of metered-dose inhaler electronic monitoring in a real-world asthma randomized controlled trial
Related Experiment Video
Updated: May 6, 2026

Conducting Respiratory Oscillometry in an Outpatient Setting
Published on: April 8, 2022
Mitesh Patel1, Janine Pilcher, Justin Travers
1Medical Research Institute of New Zealand, Wellington, New Zealand; Capital & Coast District Health Board, Wellington, New Zealand; Division of Respiratory Medicine, School of Clinical Sciences, University of Nottingham, Nottingham, United Kingdom.
Background:
Electronic monitoring of inhaled asthma medications is one method to measure medication adherence and patterns of use. Information on the performance of monitors in a randomized controlled trial allows researchers and clinicians to understand their utility and limitations. The Smartinhaler Tracker is an electronic monitor for metered-dose inhalers (MDIs) that records the date, time, and number of actuations.
Objective:
To determine the performance of the Smartinhaler monitors used in a 24-week randomized controlled trial of 303 patients with asthma in a real-world setting.
Methods:
Prestudy use checks involved 2 actuations of the MDI, with a further 2 performed 2 hours later. Within-study monitor checks, performed before dispensing at clinic visits 2 to 4, included a computerized check of monitor clock function, actuation accuracy, and battery life. Within-study data checks involved computerized checks of monitor clock function before data upload.
Results:
Two thousand six hundred seventy-eight of 2728 monitors (98.2%) passed prestudy use checks. Seventy-six of 2642 monitors (2.9%) dispensed to participants failed within-study monitor checks. Fifty-one of 2642 monitors (1.9%) malfunctioned before data upload, mostly as a result of fluid immersion. Ninety-three of 2642 monitors (3.5%) were lost or thrown away by participants. Complete data was available from 2498 of 2642 dispensed monitors (94.5%) and 2498 of 2549 returned monitors (98.0%).
Conclusions:
The Smartinhaler Tracker is a reliable monitor for measuring MDI use in a real-world setting. Use of extensive monitor and data-checking protocols reduces data loss. In a research or clinical setting, the use of a validated and reliable electronic monitor represents the reference standard for assessing patterns of medication use.
Related Concept Videos
Inhaled Medications
Asthma-IV: Diagnostic and Management
Clinical Assessment for Asthma:
This is the first step in diagnosing and managing asthma. It includes:
COPD: Management Using Bronchodilators and Corticosteroids
Antiasthma Drugs: β2-Adrenoceptor Agonists
One class of bronchodilators includes β2-adrenoceptor agonists. These agents target the β2-adrenoceptors located on bronchial smooth muscle cells. By stimulating these receptors, β2-agonists induce...
Asthma-IV: Nursing Management
First, in...