Related Experiment Video
Updated: May 5, 2026

Measurement of Ultrafast Vibrational Coherences in Polyatomic Radical Cations with Strong-Field Adiabatic Ionization
Published on: August 6, 2018
Novel tandem quadrupole-acceleration-deceleration mass spectrometer for neutralization-reionization studies
1Department of Chemistry, University of Washington, BG-10, 98195, Seattle, WA.
Abstract:
A new tandem mass spectrometer of the quadrupole-acceleration lens-deceleration. lens-quadrupole (QADQ) configuration is described. The instrument is designed for neutralization-reionization studies and consists of a 2000-u quadrupole mass analyzer as MS-I, an acceleration electrostatic lens, a series of three differentially pumped collision cells, and an electrostatic deceleration lens, energy filter, and another 2000-u quadrupole mass analyzer as MS-II. The ion optical system achieves high total ion transmission for 5-9-keV ions. Unit mass resolution in neutralization-reionization mass spectra of aromatic compounds is demonstrated. Mass, kinetic energy, and linked scans at various levels of mass resolution and sensitivity are described.
Related Concept Videos
Tandem Mass Spectrometry
Mass Analyzers: Common Types
Mass Spectrometers
Mass Analyzers: Overview
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Mass Spectrum: Interpretation