Analysis of Complex Molecules and Their Reactions on Surfaces by Means of Cluster-Induced Desorption/Ionization Mass Spectrometry
Published on: March 1, 2020
a compact high pressure ion source for high and low energy collision-induced dissociation studies of cluster ions on
J E Szulejko1, C E Hop, T B McMahon
1Department of Chemistry and Guelph-Waterloo Centre for Graduate Work in Chemistry, University of Waterloo, Waterloo, Ontario, Canada.
Abstract:
A compact, field-free high pressure ion source designed to replace, with minimum disruption, the electron impact/chemical ionization ion source of a VG Analytical ZAB-2FQ hybrid BEqQ mass spectrometer is described. This ion source may be operated at temperatures from ≈40 to 250 °C and at pressures up to 4-5 torr and, thus, is capable of producing proton-bound cluster ions up to hexamers in good yields. Examples of high energy collision-induced dissociation, low energy collision-induced dissociation, and neutralization-reionization studies of proton-bound cluster ions produced in this source are presented.