Supercontinuum spatial modulation spectroscopy: detection and noise limitations

M P McDonald1, F Vietmeyer, D Aleksiuk

  • 1Department of Chemistry and Biochemistry, University of Notre Dame, Notre Dame, Indiana 46556, USA.

Related Concept Videos

Difference from Background: Limit of Detection01:05

Difference from Background: Limit of Detection

The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...
9.0K
Raman Spectroscopy Instrumentation: Overview01:26

Raman Spectroscopy Instrumentation: Overview

A conventional Raman spectrophotometer includes a laser source, a sample holding system, a wavelength selector, and a detector.
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
1.8K
¹³C NMR: ¹H–¹³C Decoupling01:04

¹³C NMR: ¹H–¹³C Decoupling

The probability of having two carbon-13 atoms next to each other is negligible because of the low natural abundance of carbon-13. Consequently, peak splitting due to carbon-carbon spin-spin coupling is not observed in spectra. However, protons up to three sigma bonds away split the carbon signal according to the n+1 rule, resulting in complicated spectra.
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
1.7K
NMR Spectrometers: Overview01:20

NMR Spectrometers: Overview

NMR spectrometers consist of a strong magnet, a radiofrequency transmitter, and a detector attached to a computer console for recording spectra of samples containing NMR-active nuclei. In first-generation NMR instruments called continuous-wave spectrometers, the resonance frequencies of the nuclei are determined by frequency-sweep or field-sweep methods. The magnetic field strength is fixed and the rf signal is swept in the former, while the radiofrequency signal is fixed and the magnetic field...
2.0K
Super-resolution Fluorescence Microscopy01:37

Super-resolution Fluorescence Microscopy

Super-resolution fluorescence microscopy (SRFM) provides a better resolution than conventional fluorescence microscopy by reducing the point spread function (PSF). PSF is the light intensity distribution from a point that causes it to appear blurred. Due to PSF, each fluorescing point appears bigger than its actual size, and it is the PSF interference of nearby fluorophores that causes the blurred image. Various approaches to achieving higher resolution through SRFM have recently been...
12.3K
IR Spectroscopy: Molecular Vibration Overview01:24

IR Spectroscopy: Molecular Vibration Overview

When Infrared (IR) radiation passes through a covalently bonded molecule, the bonds transition from lower to higher vibrational levels. The fundamental vibrational motions that result in infrared absorption can be classified as stretching or bending vibrations.
Stretching vibrations are vibrational motions that occur along the bond line, changing the bond length or distance between two bonded atoms. They are further distinguished as symmetric or asymmetric. In symmetric stretching, the...
5.8K