Comparison of gamma (Anger) camera systems in terms of detective quantum efficiency using Monte Carlo simulation
Ida Eriksson1, Sven-Åke Starck, Magnus Båth
1aDepartment of Medical Physics, Karlstad Hospital, Karlstad bDepartment of Radiation Physics, Institute of Clinical Sciences, the Sahlgrenska Academy at University of Gothenburg cDepartment of Medical Physics and Biomedical Engineering, Sahlgrenska University Hospital, Gothenburg dDepartment of Natural Science and Biomedicine, School of Health Sciences, Jönköping University eHospital Physics, Department of Oncology, County Hospital Ryhov, Jönköping, Sweden.
Aim:
The aim of the present study was to perform an extensive evaluation of available gamma camera systems in terms of their detective quantum efficiency (DQE) and determine their dependency on relevant parameters such as collimator type, imaging depth, and energy window using the Monte Carlo technique.
Materials And Methods:
The modulation transfer function was determined from a simulated (99m)Tc point source and was combined with the system sensitivity and photon yield to obtain the DQE of the system. The simulations were performed for different imaging depths in a water phantom for 13 gamma camera systems from four manufacturers.
Results:
Except at very low spatial frequencies, the highest DQE values were found with a lower energy window threshold of around 130 keV for all systems. The height and shape of the DQE curves were affected by the collimator design and the intrinsic properties of the gamma camera systems. High-sensitivity collimators gave the highest DQE at low spatial frequencies, whereas the high-resolution and ultrahigh-resolution collimators showed higher DQE values at higher frequencies. The intrinsic resolution of the system mainly affected the DQE curve at superficial depths.
Conclusion:
The results indicate that the manufacturers have succeeded differently in their attempts to design a system constituting an optimal compromise between sensitivity and spatial resolution.
More Related Videos
06:08Time-resolved Photophysical Characterization of Triplet-harvesting Organic Compounds at an Oxygen-free Environment Using an iCCD Camera
Published on: December 27, 2018
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
High-Performance Liquid Chromatography: Types of Detectors
Determination of Crystal Structures
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...