Cardiac magnetic resonance before and after percutaneous pulmonary valve implantation
Francesco Secchi1, Elda Chiara Resta, Luciane Piazza
1Unità di Radiologia, IRCCS Policlinico San Donato, Piazza E. Malan, 20097, San Donato Milanese, Milan, Italy, francescosecchimd@gmail.com.
Purpose:
To measure the magnetic resonance (MR) artefact produced by a percutaneous pulmonary valve stent and to evaluate the changes in volumetric and functional right ventricle (RV) parameters due to percutaneous pulmonary valve implantation (PPVI).
Materials And Methods:
A Melody valve was studied in vitro using clinical cardiac MR (CMR) sequences. In vivo, we analysed the CMR examinations obtained before and after PPVI of 27 consecutive patients. The echocardiography pressure gradient (PG) and catheter PG were measured. The Wilcoxon test was used for comparisons.
Results:
In vitro, the least difference between artefact extent and actual valve size (0.1 mm) was obtained with a steady-state free precession (SSFP) sequence. In vivo, RV end-diastolic volume (ml/m(2)), end-systolic volume (ml/m(2)) and ejection fraction (%) were 79 ± 42, 43 ± 41 and 49 ± 13 before PPVI and 64 ± 21 (p = 0.054), 30 ± 14 (p = 0.021), and 54 ± 12 (p = 0.018) 6 months after PPVI, respectively. The PG and regurgitation fraction (RF) were 36 ± 15 mmHg and 14 ± 18 % before PPVI and 13 ± 15 mmHg (p < 0.001) and 2 ± 5 % (p = 0.013) after PPVI, respectively. No significant differences were found comparing the PG measured with CMR, echocardiography and catheter.
Conclusions:
We showed in vitro that the SSFP sequence produced the most accurate valve measurement. After PPVI, CMR showed a strong decrease of PG and RF with a significant improvement of RV function.
More Related Videos
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
09:57Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
Related Concept Videos
Mitral Stenosis II: Clinical features and Diagnostic Tests
Mitral Valve Prolapse II: Assessment and Management
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...