Alternative representation for the stability diagram of quadrupole ion traps upon additional quadrupolar excitation
Arnaud Vernier1, Fabien Chirot, Rodolphe Antoine
1Université de Lyon, UMR5306, Université Lyon 1, CNRS, France.
Abstract:
We present a combined theoretical and experimental study of the stability of ions in a linear ion trap under the application of one or two auxiliary radiofrequency (RF) fields, in order to perform simultaneous resonant excitation/ejection of several different ions. The influence of the amplitude and frequency of the auxiliary field is addressed through the construction of experimental and theoretical stability diagrams. Theoretical diagrams are constructed using the method developed by Konenkov et al. [J. Am. Soc. Mass Spectrom. 13, 597 (2002)]. We propose a new representation of stability diagrams more adapted to the study of auxiliary excitations than the canonical one. Stability regions are represented as a function of the fundamental RF amplitude and of the relative intensity of the excitation. This representation facilitates the monitoring of the evolution of the mass-selectivity of first- and higher-order resonant excitations in the trap, for which an empirical law is derived. We also show that the relative phase shift between the excitation field and the main driving field has a strong influence on the shape of the diagrams.
Related Concept Videos
Mass Analyzers: Common Types
Atomic Nuclei: Nuclear Relaxation Processes
¹H NMR: Complex Splitting
Splitting diagrams or splitting tree diagrams are routinely used to depict such complex couplings. While drawing splitting diagrams, the splitting with the larger coupling constant is usually applied...
Interpreting ¹H NMR Signal Splitting: The (n + 1) Rule
Deactivation Processes: Jablonski Diagram
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...