Related Experiment Video
Updated: May 5, 2026

Effects of Mechanical Methods Used in Peri-implantitis Treatment on Implant Surface Decontamination and Roughness
Published on: March 14, 2025
Accuracy of electronic apex locator in enlarged root canals with different root canal irrigants: an in vitro study
Sainath Dinapadu1, Srikanth Pasari, Shilpa Reddy Admala
1Senior Lecturer, Department of Conservative Dentistry and Endodontics, SVS Institute of Dental Sciences, Mahabubnagar, Andhra Pradesh, India, Phone: 91-8542-271515, Fax: 91-8542-273111,
Aim:
To evaluate the accuracy of Root ZX-II (J Morita Corp) apex locator in enlarged root canals with different root canal irrigants.
Materials And Methods:
48 freshly extracted single rooted mandibular premolar teeth were used. The apical enlargement was done up to #45 K-file as the master apical file (MAF). The teeth were randomly divided into 4 groups and mounted in an experimental apparatus. The following irrigants were used during electronic canal measurements: group 1: saline; group 2: 3% NaOCl; group 3: 2% chlorhexidine; group 4: 17% EDTA. The canal measurement was done with Root ZX-II apex locator using #10 and #45 K-file. Data were analyzed by two way ANOVA and Bonferroni multiple comparison test.
Results:
Statistical analysis showed that there was statistically significant difference between #10 and #45 K-file for group 1 (p < 0.01).
Conclusion:
Root ZX-II was accurate in the presence of 3% NaOCl and 17% EDTA when measured with smaller and larger files. However, it was accurate in the presence of saline and 2% CHX when larger files were used.