Arbitrary magnetic field gradient waveform correction using an impulse response based pre-equalization technique

Frédéric G Goora1, Bruce G Colpitts2, Bruce J Balcom3

  • 1Department of Electrical and Computer Engineering, University of New Brunswick, 15 Dineen Drive, Fredericton, NB E3B 5A3, Canada; MRI Centre, Department of Physics, University of New Brunswick, 8 Bailey Drive, Fredericton, NB E3B 5A3, Canada.

Related Concept Videos

NMR Spectrometers: Resolution and Error Correction01:14

NMR Spectrometers: Resolution and Error Correction

When magnetic nuclei in a sample achieve resonance and undergo relaxation, the signal detected in NMR is an approximately exponential free induction decay. Fourier transform of an exponential decay yields a Lorentzian peak in the frequency domain. Lorentzian peaks in an NMR spectrum are defined by their amplitude, full width at half maximum, and position, where the peak width is governed by the spin-spin relaxation time alone. In real experiments, however, the applied magnetic field is rendered...
993
Plane Electromagnetic Waves II01:29

Plane Electromagnetic Waves II

Consider a plane wavefront traveling in position x-direction with a constant speed. This wavefront can be utilized to obtain the relationship between electric and magnetic fields with the help of Faraday's law.
3.1K
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences01:17

NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences

A pulse is a short burst of radio waves distributed over a range of frequencies that simultaneously excites all the nuclei in the sample. Upon passing a radio frequency pulse along the x-axis, the nuclei absorb energy corresponding to their Larmor frequencies and achieve resonance. This shifts the net magnetization vector from the z-axis toward the transverse plane. This angle of rotation of the magnetization vector, or the flip angle, is proportional to the duration and intensity of the pulse.
1.8K
Atomic Nuclei: Larmor Precession Frequency01:11

Atomic Nuclei: Larmor Precession Frequency

The earth's gravitational field produces a 'twisting force' perpendicular to the angular momentum of a spinning mass (such as a spinning top) that causes the mass to 'wobble' around the gravitational field axis in a phenomenon called precession. Similarly, the magnetic moment (μ) of a spinning nucleus precesses due to an external magnetic field directed along the z-axis. The precession of the magnetic moment vector about the magnetic field is called Larmor precession,...
3.5K
Reconstruction of Signal using Interpolation01:10

Reconstruction of Signal using Interpolation

Signal processing techniques are essential for accurately converting continuous signals to digital formats and vice versa. When a continuous signal is sampled with a period T, the resulting sampled signal exhibits replicas of the original spectrum in the frequency domain, spaced at intervals equal to the sampling frequency. To handle this sampled signal, a zero-order hold method can be applied, which creates a piecewise constant signal by retaining each sample's value until the next...
920
Magnetic Field Of A Current Loop01:16

Magnetic Field Of A Current Loop

Consider a circular loop with a radius a, that carries a current I. The magnetic field due to the current at an arbitrary point P along the axis of the loop can be calculated using the Biot-Savart law.
6.1K