Related Experiment Video
Updated: May 5, 2026

Low-Dose Gamma Radiation Sterilization for Decellularized Tracheal Grafts
Published on: April 14, 2023
Sterilization by gamma radiation of antibiotic impregnated polymethylmethacrylate and plaster of Paris beads. A pilot
P Trencart, Y A Elce1, E Rodriguez Batista
1Yvonne A. Elce, DVM DACVS, Department of Clinical Sciences, Faculty of Veterinary Medicine, University of Montreal, St Hyacinthe, Quebec, J2S 7C6, Canada, Phone: +1 450 778 8100,
Introduction:
Ethylene oxide is currently recommended for sterilization of antibiotic impregnated beads; however this method carries health risks to personnel and is becoming less available.
Objective:
To perform a pilot study of the effect of radiation for sterilization of polymethylmethacrylate (PMMA) and plaster of Paris (POP) beads impregnated with amikacin, enrofloxacin, and ceftiofur.
Hypothesis:
Radiation would effectively sterilize the beads without affecting the efficacy of the antibiotic.
Materials And Methods:
Beads of PMMA and POP were prepared in a clean but non-sterile manner with one of the three antibiotics (amikacin, enrofloxacin, ceftiofur) or no antibiotic. Beads were then exposed to radiation for a total dose of 0 kiloGray (kGy), 10 kGy and 25 kGy. Beads were incubated on Mueller-Hinton agar plates seeded with Escherichia coli, Staphylococcus aureus or Pseudomonas aeruginosa for 24 hours or cultured in brain-heart infusion broth for 48 hours. Zones of inhibition were measured on the agar plates and statistics were performed on the diameters of the zones of inhibition using an analysis of variance.
Results:
There were no differences in the diameters of inhibition for all levels of radiation for all PMMA beads. The same was true with POP beads with the exception of enrofloxacin which had a significantly decreased zone of inhibition with increased levels of radiation, though the clinical significance of this finding was not assessed. Only beads without antibiotics and not exposed to radiation had bacterial growth.
Clinical Significance:
Radiation may be an effective method of sterilization for antibiotic impregnated beads.
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration