Related Experiment Video
Updated: May 5, 2026

Biaxial Mechanical Characterizations of Atrioventricular Heart Valves
Published on: April 9, 2019
Determination of the tensile mechanical properties of the segmented mitral valve annulus
Gillian M Gunning1, Bruce P Murphy1
1Trinity Centre for Bioengineering, Trinity Biomedical Sciences Institute, Trinity College, 152-160 Pearse Street, Trinity College Dublin, Dublin 2, Ireland; Department of Mechanical and Manufacturing Engineering, School of Engineering, Trinity College Dublin, Dublin 2, Ireland.
Abstract:
The mitral valve annulus is a complex and irregular component of the mitral valve apparatus, serving both a structural and sphincteric role. We have sought to determine the mechanical properties of the mitral valve annulus segmentally. Twenty porcine hearts were dissected to isolate the annulus. The annulus was segmented into four sections: anterior, posterior, and left and right commissural sections. Ten of these were tensile tested to failure as control samples. The remaining ten were digested in order to fully isolate the annulus from the myocardium, and subsequently tensile tested to failure. Histological samples of each segment were analysed to determine collagen/annular content. Whole segments of muscular annulus were tensile tested to failure; the stress and strain at failure and location of failure were determined in these larger specimens. Our results demonstrated that the anterior annulus is stiffer than the posterior segment by a factor of approximately 27 at a 2% strain level, and approximately 13 at a 6% strain. There is a trend in the results that identifies that the muscular annulus is stiffest at the right commissural segment, while the posterior segment tends to be the least stiff. The stiffness of the samples can be correlated with the area associated with the dense collagen annulus using histological analysis. Finally, the weakest section of the mitral valve annulus was identified as the intersection of the right commissural segment and the posterior segment.
More Related Videos
07:34Layer Microdissection of Tricuspid Valve Leaflets for Biaxial Mechanical Characterization and Microstructural Quantification
Published on: February 10, 2022
11:38Environmentally-controlled Microtensile Testing of Mechanically-adaptive Polymer Nanocomposites for ex vivo Characterization
Published on: August 20, 2013
Related Concept Videos
Mechanical Characteristics of Steel
The tension test is fundamental for determining tensile strength. In this test, a steel specimen is stretched using a gripping device until it breaks. The data collected during this test are used...
Mitral Valve Prolapse I: Introduction