Low concentration contrast medium for dual-source computed tomography coronary angiography by a combination of
Minwen Zheng1, Ying Liu1, Mengqi Wei1
1Department of Radiology, Xijing Hospital, Fourth Military Medical University, 127# West Changle Road, Xi'an 710032, Shaanxi Province, China.
Objectives:
To assess the impact of low-concentration contrast medium on vascular enhancement, image quality and radiation dose of coronary CT angiography (cCTA) by using a combination of iterative reconstruction (IR) and low-tube-voltage technique.
Materials And Methods:
One hundred patients were prospectively randomized to two types of contrast medium and underwent prospective electrocardiogram-triggering cCTA (Definition Flash, Siemens Healthcare; collimation: 128 mm × 0.6mm; tube current: 300 mAs). Fifty patients received Iopromide 370 were scanned using the conventional tube setting (100 kVp or 120 kVp if BMI ≥ 25 kg/m(2)) and reconstructed with filtered back projection (FBP). Fifty patients received Iodixanol 270 were scanned using the low-tube-voltage (80 kVp or 100 kVp if BMI ≥ 25 kg/m(2)) technique and reconstructed with IR. CT attenuation was measured in coronary artery and other anatomical regions. Noise, image quality and radiation dose were compared.
Results:
Compared with two Iopromide 370 subgroups, Iomeprol 270 subgroups showed no significant difference in CT attenuation (576.63 ± 95.50 vs. 569.51 ± 118.93 for BMI< 25 kg/m(2), p=0.647 and 394.19 ± 68.09 vs. 383.72 ± 63.11 for BMI ≥ 25 kg/m(2), p=0.212), noise (in various anatomical regions of interest) and image quality (3.5 vs. 4.0, p=0.13), but significantly (0.41 ± 0.17 vs. 0.94 ± 0.45 for BMI< 25 kg/m(2), p<0.001 and 1.14 ± 0.24 vs. 2.37 ± 0.69 for BMI ≥ 25 kg/m(2), p<0.001) lower radiation dose, which reflects dose saving of 56.4% and 51.9%, respectively.
Conclusions:
Combined IR with low-tube-voltage technique, a low-concentration contrast medium of 270 mg I/ml can still maintain the contrast enhancement without impairing image quality, as well as significantly lower the radiation dose.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT