Related Experiment Video
Updated: May 4, 2026

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
Whole-brain three-dimensional T2-weighted BOLD functional magnetic resonance imaging at 7 Tesla
Jun Hua1, Qin Qin, Peter C M van Zijl
1The Russell H. Morgan Department of Radiology and Radiological Science, Division of MR Research, The Johns Hopkins University School of Medicine, Baltimore, Maryland, USA; F.M. Kirby Research Center for Functional Brain Imaging, Kennedy Krieger Institute, Baltimore, Maryland, USA.
Purpose:
A new acquisition scheme for T2-weighted spin-echo BOLD fMRI is introduced.
Methods:
It uses a T2-preparation module to induce blood-oxygenation-level-dependent (BOLD) contrast, followed by a single-shot three-dimensional (3D) fast gradient-echo readout with short echo time (TE). It differs from most spin-echo BOLD sequences in that BOLD contrast is generated before the readout, which eliminates the "dead time" due to long TE required for T2 contrast, and substantially improves acquisition efficiency. This approach, termed "3D T2prep-GRE," was implemented at 7 Tesla (T) with a typical spatial (2.5 × 2.5 × 2.5 mm(3) ) and temporal (TR = 2.3 s) resolution for functional MRI (fMRI) and whole-brain coverage (55 slices), and compared with the widely used 2D spin-echo EPI sequence.
Results:
In fMRI experiments of simultaneous visual/motor activities, 3D T2prep-GRE showed minimal distortion and little signal dropout across the whole brain. Its lower power deposition allowed greater spatial coverage (55 versus 17 slices with identical TR, resolution and power level), temporal SNR (60% higher) and CNR (35% higher) efficiency than 2D spin-echo EPI. It also showed smaller T2* contamination.
Conclusion:
This approach is expected to be useful for ultra-high field fMRI, especially for regions near air cavities. The concept of using T2-preparation to generate BOLD contrast can be combined with many other sequences at any field strength.
Related Concept Videos
Magnetic Resonance Imaging
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...
Imaging Studies IV: Magnetic Resonance Imaging