Correction of proton resonance frequency shift MR-thermometry errors caused by heat-induced magnetic susceptibility
Paul Baron1, Roel Deckers, Martijn de Greef
1Image Sciences Institute, University Medical Center Utrecht, Utrecht, The Netherlands.
Purpose:
In this study, we aim to demonstrate the sensitivity of proton resonance frequency shift (PRFS) -based thermometry to heat-induced magnetic susceptibility changes and to present and evaluate a model-based correction procedure.
Theory And Methods:
To demonstrate the expected temperature effect, field disturbances during high intensity focused ultrasound sonications were monitored in breast fat samples with a three-dimensional (3D) gradient echo sequence. To evaluate the correction procedure, the interface of tissue-mimicking ethylene glycol gel and fat was sonicated. During sonication, the temperature was monitored with a 2D dual flip angle multi-echo gradient echo sequence, allowing for PRFS-based relative and referenced temperature measurements in the gel and T1 -based temperature measurements in fat. The PRFS-based measurement in the gel was corrected by minimizing the discrepancy between the observed 2D temperature profile and the profile predicted by a 3D thermal model.
Results:
The HIFU sonications of breast fat resulted in a magnetic field disturbance which completely disappeared after cooling. For the correction method, the 5th to 95th percentile interval of the PRFS-thermometry error in the gel decreased from 3.8°C before correction to 2.0-2.3°C after correction.
Conclusion:
This study has shown the effects of magnetic susceptibility changes induced by heating of breast fatty tissue samples. The resultant errors can be reduced by the use of a model-based correction procedure.
More Related Videos
07:38Real-time Monitoring of High Intensity Focused Ultrasound HIFU Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound HMIFU
Published on: November 3, 2015
08:08Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Published on: March 6, 2019
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...