A TPD and AR based comparison of accelerator neutron irradiation fields between (7)Li and W targets for BNCT
Kenichi Tanaka1, Satoru Endo2, Shunsuke Yonai3
1Center of Medical Education, Sapporo Medical University, Sapporo, Japan.
Abstract:
The characteristics of moderator assembly dimension was investigated for the usage of (7)Li(p,n) neutrons by 2.3-2.8MeV protons and W(p,n) neutrons by 50MeV protons. The indexes were the treatable protocol depth (TPD) and advantage depth (AD). Consequently, a configuration for W target with the Fe filter, Fluental moderator, Pb reflector showed the TPD of 5.8cm and AD of 9.3cm. Comparable indexes were found for the Li target in a geometry with the MgF2 moderator and Teflon reflector.