Double sensing by atrial automatic tachycardia-terminating pulse generator.

S S Barold, M D Falkoff, L S Ong

    Pacing and Clinical Electrophysiology : PACE
    |January 1, 1987
    PubMed
    Summary

    Double sensing of atrial and QRS signals by an implanted pulse generator caused inappropriate treatment for supraventricular tachycardia. The far-field QRS signal is crucial for accurate tachycardia detection in automatic pulse generators.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Remaining agile - post-pandemic prevention of SARS-CoV-2 nosocomial transmission.

    The Journal of hospital infection·2023
    Same author

    True vs. pseudo-electrical ventricular alternans sustained by ventricular premature complexes.

    Herzschrittmachertherapie & Elektrophysiologie·2018
    Same author

    A review of the atrial upper rate algorithms of St. Jude Medical (Abbott) cardiac implantable electronic devices : Incidence of repetitive nonreentrant ventriculoatrial synchrony (RNRVAS).

    Herzschrittmachertherapie & Elektrophysiologie·2017
    Same author

    Erratum to: Noise from a dysfunctional atrial lead detected as atrial fibrillation by a cardiac implantable electronic device.

    Herzschrittmachertherapie & Elektrophysiologie·2017
    Same author

    Diagnosis of myocardial infarction and ischemia in the setting of bundle branch block and cardiac pacing.

    Herzschrittmachertherapie & Elektrophysiologie·2016
    Same author

    Unusual cause of desynchronization in a cardiac resynchronization device.

    Herzschrittmachertherapie & Elektrophysiologie·2012

    Area of Science:

    • Cardiology
    • Biomedical Engineering
    • Medical Devices

    Background:

    • Supraventricular tachycardia (SVT) is a common arrhythmia requiring therapeutic intervention.
    • Implantable automatic burst tachycardia-terminating pulse generators are used for SVT treatment.
    • Accurate sensing of cardiac signals is critical for device efficacy and patient safety.

    Observation:

    • Two cases of double sensing of atrial (P) and ventricular (QRS) signals were observed.
    • This occurred with an Intermedics CyberTach 60 implanted pulse generator during normal sinus rhythm.
    • The device inappropriately delivered burst stimulation due to sensing at rates slower than programmed detection criteria.

    Findings:

    • The far-field QRS signal interfered with the atrial electrogram sensing.

    Related Experiment Videos

  • This interference led to misinterpretation of normal sinus rhythm as SVT.
  • Inappropriate burst stimulation was delivered, highlighting a sensing malfunction.
  • Implications:

    • The far-field QRS signal plays a significant role in SVT detection by atrial sensing pulse generators.
    • Device programming and sensing parameters require careful consideration to prevent double sensing.
    • Further research into advanced signal processing algorithms for implantable devices is warranted to improve accuracy and prevent inappropriate therapies.