Stent longitudinal strength assessed using point compression: insights from a second-generation, clinically related
John A Ormiston1, Bruce Webber, Ben Ubod
1From the Mercy Angiography, Auckland, New Zealand (J.A.O., B.W., B.U., M.W.I.W.); and Cardiology Department, Auckland City Hospital, Auckland, New Zealand (J.A.O., J.W., M.W.I.W.).
This study tested stent distortion resistance in a new, clinically relevant model. The Biomatrix Flex stent showed the most resistance, while the Element stent was most susceptible to deformation.
Area of Science:
- Biomaterials Engineering
- Medical Device Design
- Cardiovascular Research
Background:
- Stent longitudinal distortion, though rare, can cause adverse clinical events.
- Previous tests applied force to the entire stent hoop.
- A new test simulates clinical scenarios by applying force to a malapposed stent hoop.
Purpose of the Study:
- To evaluate and compare the susceptibility of different stent designs to longitudinal distortion.
- To develop a more clinically relevant bench test for stent deformation.
- To provide insights into stent distortion mechanisms in real-world applications.
Main Methods:
- Six stent designs (5 examples each) with 3-mm diameters were tested.
- Stents were malapposed in a 3.5-mm tube, with the proximal 5 mm intentionally displaced.
- An Instron device applied localized force to the proximal stent hoop, measuring force and compression.
- Deformation, malapposition, and lumen obstruction were assessed.
Main Results:
- The Biomatrix Flex stent exhibited the highest resistance to deformation.
- The Element stent demonstrated the greatest susceptibility to distortion.
- Promus Premier design modifications reduced distortion potential compared to the Element.
- Vision, Multi-Link 8, and Integrity designs showed moderate distortion resistance.
Conclusions:
- The second-generation test provides a more accurate assessment of clinical stent distortion.
- Stent design significantly impacts resistance to longitudinal distortion.
- Specific design features, like increased connectors, can improve distortion resistance.
More Related Videos
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
07:09In situ Compressive Loading and Correlative Noninvasive Imaging of the Bone-periodontal Ligament-tooth Fibrous Joint
Published on: March 7, 2014
Related Concept Videos
Mechanical Characteristics of Steel
The tension test is fundamental for determining tensile strength. In this test, a steel specimen is stretched using a gripping device until it breaks. The data collected during this test are used...
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in...
Yield Criteria for Ductile Materials under Plane Stress
The Maximum Shearing Stress Criterion, also known as...