Objective:
To simulate the multi-leaf collimator of Varian linear accelerator using Monte Carlo method.
Methods:
The multi-leaf collimator model was established using the DYNVMLC module of BEAMnrc and validated by comparison of Monte Carlo simulation and actual measurement results.
Results:
The simulation results were well consistent with the actual measurement results with a bias of less than 3%.
Conclusion:
The multi-leaf collimator of Varian linear accelerator can be successfully modeled using Monte Carlo method for analysis of the impact of the geometric properties of the multi-leaf collimator on the dose distribution.
More Related Videos
08:25Radiation Planning Assistant - A Streamlined, Fully Automated Radiotherapy Treatment Planning System
Published on: April 11, 2018
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Radiation: Applications
The average...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...