Related Experiment Video
Updated: May 4, 2026

Diagonal Method to Measure Synergy Among Any Number of Drugs
Published on: June 21, 2018
Bidirectional (negative/positive) interference of oleandrin and oleander extract on a relatively new Loci digoxin
Amitava Dasgupta1, Kerry J Welsh, Shen-An Hwang
1Department of Pathology and Laboratory Medicine, The University of Texas Medical School at Houston, Texas.
Background:
Oleander interferes with serum digoxin measurements using various immunoassays. The potential interference of oleander and its active ingredient, oleandrin, with a relatively new homogenous sequential chemiluminescent digoxin assay based on luminescent oxygen channeling technology (LOCI digoxin assay, Siemens Diagnostics) has not been previously reported.
Methods:
Aliquots of a digoxin-free serum pool were supplemented with increasing concentrations of oleandrin, or with oleander extract, followed by measuring the apparent digoxin concentrations using the LOCI digoxin assay using Vista 1500 analyzer. Mice were fed oleandrin or oleander extract, and their blood digoxin levels at 1 and 2 h were measured with the LOCI digoxin assay. In addition, two digoxin serum pools were prepared by combining sera of patients receiving digoxin; aliquots of both pools were supplemented with oleandrin or oleander extract and digoxin concentrations were again measured. Attempts to overcome this interference were made by measuring free digoxin concentration using a third digoxin pool.
Results:
Significant apparent digoxin concentrations were observed after supplementing aliquots of the drug-free serum pool with oleandrin or oleander extract. Mice fed with oleandrin or oleander extract also showed apparent digoxin levels 1 and 2 h after feeding. Digoxin values were also falsely lower or elevated (bidirectional interference) when aliquots of digoxin serum pools were further supplemented with oleandrin or oleander extract depending on concentration; this interference was not eliminated by free digoxin monitoring.
Conclusions:
Oleandrin interferes with LOCI digoxin assay.
More Related Videos
10:41Laser-Induced Action Potential-Like Measurements of Cardiomyocytes on Microelectrode Arrays for Increased Predictivity of Safety Pharmacology
Published on: September 13, 2022
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Related Concept Videos
Heart Failure Drugs: Inotropic Agents
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Therapeutic Drug Monitoring: Drug Analysis Methods
Therapeutic Drug Monitoring: Affecting Factors