Beam rate influence on dose distribution and fluence map in IMRT dynamic technique
Krzysztof Slosarek1, Aleksandra Grządziel1, Wojciech Osewski1
1Radiotherapy and Brachytherapy Planning Department, Center of Oncology - MSC Memorial Institute, Gliwice Branch, Poland.
Lower beam rates in intensity-modulated radiation therapy (IMRT) planning offer slightly better dose distribution and accuracy. While beam rate affects dynamic techniques, the impact on overall treatment accuracy is minimal, favoring slower rates.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Technology
Background:
- Radiotherapy accelerators use variable beam rates to shorten treatment times and minimize patient motion.
- This optimization is crucial for advanced techniques like intensity-modulated radiation therapy (IMRT).
Purpose of the Study:
- To investigate the effect of different beam rates on dose distribution in IMRT plans.
- To assess the agreement between calculated and measured dose distributions across various beam rates.
Main Methods:
- IMRT plans were analyzed at two beam rates: 100 MU/min and 600 MU/min.
- Comparisons included Radiation Planning Index (RPI), monitor units (MU), and fluence map accuracy.
- Measured dose distributions were evaluated against calculated ones using the Gamma index.
Main Results:
- Lower beam rates (100 MU/min) yielded slightly improved target coverage and organ-at-risk protection.
- Plans at 100 MU/min required fewer MUs and showed greater similarity between optimal and actual fluence maps.
- Measured dose distributions more closely matched calculated ones at lower beam rates, though differences were minor.
Conclusions:
- Beam rate influences dynamic IMRT execution but results in minor differences in dose distribution.
- A slight superiority of lower beam rates was observed, without significant impact on overall treatment accuracy.