Identification of a unique cause of ring artifact seen in computed tomography trans-axial images
Ashish Kumar Jha1, Nilendu C Purandare1, Sneha Shah1
1Department of Nuclear Medicine and Molecular Imaging, Tata Memorial Hospital, Parel, Mumbai, Maharashtra, India.
Abstract:
Artifacts present in computed tomography (CT) image often degrade the image quality and ultimately, the diagnostic outcome. Ring artifact in trans-axial image is caused by either miscalibrated or defective detector element of detector row, which is often categorized as scanner based artifact. A ring artifact detected on trans-axial CT image of positron emission tomography/computed tomography (PET/CT), was caused by contamination of CT tube aperture by droplet of injectable contrast medium. This artifact was corrected by removal of contrast droplet from CT tube aperture. The ring artifact is a very common artifact, commonly cited in the literature. Our case puts forward an uncommon cause of this artifact and its method of correction, which also, has no mention in the existing literature.
More Related Videos
07:06Use of Micro X-ray Computed Tomography with Phosphotungstic Acid Preparation to Visualize Human Fibromuscular Tissue
Published on: September 5, 2019
08:39Longitudinal Micro-Computed Tomography Image Analysis for User-Defined Region of Interest in Critical-Sized Bone Defects
Published on: June 24, 2025
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography