Low contrast volume run-off CT angiography with optimized scan time based on double-level test bolus
Jan Baxa1, Tomáš Vendiš1, Jiří Moláček2
1Department of Imaging Methods, University Hospital and Charles University Medical School in Pilsen, Czech Republic.
Purpose:
To verify the technical feasibility of low contrast volume (40 mL) run-off CT angiography (run-off CTA) with the individual scan time optimization based on double-level test bolus technique.
Materials And Methods:
A prospective study of 92 consecutive patients who underwent run-off CTA performed with 40 mL of contrast medium (injection rate of 6 mL/s) and optimized scan times on a second generation of dual-source CT. Individual optimized scan times were calculated from aortopopliteal transit times obtained on the basis of double-level test bolus technique--the single injection of 10 mL test bolus and dynamic acquisitions in two levels (abdominal aorta and popliteal arteries). Intraluminal attenuation (HU) was measured in 6 levels (aorta, iliac, femoral and popliteal arteries, middle and distal lower-legs) and subjective quality (3-point score) was assessed. Relations of image quality, test bolus parameters and arterial circulation involvement were analyzed.
Results:
High mean attenuation (HU) values (468; 437; 442; 440; 342; 274) and quality score in all monitored levels was achieved. In 91 patients (0.99) the sufficient diagnostic quality (score 1-2) in aorta, iliac and femoral arteries was determined. A total of 6 patients (0.07) were not evaluable in distal lower-legs. Only the weak indirect correlation of image quality and test-bolus parameters was proved in iliac, femoral and popliteal levels (r values: -0.263, -0.298 and -0.254). The statistically significant difference of the test-bolus parameters and image quality was proved in patients with occlusive and aneurysmal disease.
Conclusion:
We proved the technical feasibility and sufficient quality of run-off CTA with low volume of contrast medium and optimized scan time according to aortopopliteal transit time calculated from double-level test bolus.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...