Performance evaluation of the automated morphological analysis of erythrocytes by CellaVision DM96
Imiko Maenou1, Yoko Tabe2, Hans-Inge Bengtsson3
1Division of Clinical Laboratory, Juntendo University Hospital, Bunkyo-ku, Tokyo, Japan.
Background:
Automated digital morphology systems are utilized for blood cell morphological examination. The aim of this study is to evaluate the accuracy and efficacy of RBC morphological anomaly screening using the CellaVision DM96 (DM96) automated image analysis system.
Methods:
The automated analysis of RBC shape, size, and chromasia abnormalities was conducted on the DM96 using 478 blood samples. A manual microscopic review was independently performed.
Results:
The DM96 preclassified samples as poikilocytosis-positive for 98% of cases with schistocytosis or echinocytosis, 97% of elliptocytosis, and 92% or 65% of cases that were positive for teardrop cells or for target cells, respectively. The accuracy of the DM96 in the detection of RBC size and chromasia abnormalities of iron deficiency anemia cases was higher than direct microscopic observation.
Conclusions:
Automated morphological analysis with the DM96 has potential utility in the morphological screening of RBC anomalies that are associated with disease.