Preliminary results: prospective clinical study to assess image-based iterative reconstruction for abdominal computed
Sarvenaz Pourjabbar1, Sarabjeet Singh, Anand K Singh
1From the *Department of Radiology, Massachusetts General Hospital, Boston, MA; and †Center for Medical Image Science and Visualization, Linköpings universitet/US, Linköping, Sweden.
Objective:
The objective of this study was to compare image quality for abdominal computed tomographic (CT) images acquired at 200 and 50 mA s and reconstructed with image-based iterative reconstruction.
Materials And Methods:
In this institutional review board-approved prospective study, 22 patients (mean [SD] age, 64.3 [14.4] years; male-female ratio, 12:10) gave informed consent for acquisition of additional abdominal CT images on 64-slice multi-detector CT (MDCT) (Siemens Definition Flash). Standard-dose images were acquired at 200 quality reference mA s, whereas low-dose images were acquired at 50 mA s (all series: 120 kV; 5-mm section thickness; pitch, 0.9:1). The low-dose images were reconstructed with a nonlinear 3-dimensional iterative image reconstruction (3D-IIR) (SafeCT; MedicVision, Tirat Carmel, Israel) (4 settings, namely, A1, A2, A3, and A4) and were assessed by 3 abdominal radiologists for lesion detection, image noise, and visibility of small structures. CATPHAN 500 was scanned at the respective doses to obtain noise spectral density and modulation transfer function.
Results:
Subjective image noise was unacceptable at 50-mA s filtered back projection and improved to average in 50-mA s A1 and minimal or no noise in 50-mA s A4. However, the visibility of small structures was similar to standard-dose filtered back projection images on 50-mA s A2. Objective image noise was reduced to 66% for the 50-mA s 3D-IIR images (9.08 [2.3]/26.75 [6.8]). The modulation transfer function curve demonstrated resolution improvement in the low-dose images with the 3D-IIR technique, whereas the noise spectral density curve confirmed noise suppression in the 50-mA s 3D-IIR images.
Conclusions:
Three-dimensional iterative image reconstruction helps to lower image noise without affecting the visibility of small structures at "moderate" settings. Diagnostically acceptable abdominal CT examinations can be acquired at 75% lower-radiation dose with the help of the image-based iterative reconstruction technique.
More Related Videos
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET